‘Undercover’ ICDs Broaden Options for Ventricular Tachyarrhythmias

Subcutaneous ICDs are better suited to some patients than their transvenous counterparts

BostonScientificSICDTorso_ICD-690

Subcutaneous implantable cardiac defibrillators (ICDs) are a less-invasive and potentially better alternative to transvenous ICDs for some patients with life-threatening ventricular tachyarrhythmias, particularly those on dialysis or who have infections or venous access issues.

Advertisement

Cleveland Clinic is a non-profit academic medical center. Advertising on our site helps support our mission. We do not endorse non-Cleveland Clinic products or services. Policy

“Implant technology has changed a lot over the years, and so have the data supporting it,” says Bruce Wilkoff, MD, Director of Cardiac Pacing and Tachyarrhythmia Devices in Cleveland Clinic’s Robert and Suzanne Tomsich Department of Cardiovascular Medicine.

Transvenous vs. subcutaneous

Because a subcutaneous ICD lead does not enter the vascular space and is not in direct contact with the heart, arrhythmia detection is accomplished differently than with transvenous ICDs.

Transvenous ICD leads fed directly into the vascular space detect heart rate only from a specific location dictated by the leads’ placement. This method provides limited insight about electrical activity outside the target area. Although it allows for excellent sensitivity in the detection of ventricular tachyarrhythmias, the trade-off can be diminished specificity and frequent inappropriate shocks. Performance is highly dependent on the programming of the device.

“With the subcutaneous ICD, you are looking at a larger view of the heart because the device does not actually touch the heart and performance is less dependent on the programming,” says Dr. Wilkoff.

How arrhythmia detection differs with subcutaneous ICDs

By using a series of sensing electrodes to record vectors of cardiac electrical conduction and the application of discrimination algorithms, subcutaneous ICDs “wait” until there is certainty that a tachyarrhythmia is detected before shocking a patient’s heart back into rhythm.

“This is a brand-new way of identifying arrhythmias,” Dr. Wilkoff says. “This algorithmic approach may very well be a superior way of going about it.”

Advertisement

Subcutaneous placement obviates the possibility of leads dislodging from the heart. In addition, implanting and explanting electrodes is much easier with a subcutaneous ICD.

MADIT-RIT findings bolster subcutaneous approach

A recent randomized trial showed that patients with tachyarrhythmias who have implantable ICDs fare better when the devices are programmed to wait out irregular heartbeats that aren’t sustained.

The study, known as MADIT-RIT (Multicenter Automatic Defibrillator Implantation Trial — Reduce Inappropriate Therapy), demonstrated that ICD therapy results in fewer inappropriate shocks and reduced all-cause mortality when the device is programmed to detect ventricular tachycardia or fibrillation with two strategies: (1) delivering therapy at a heart rate of 200 bpm or higher and (2) using a prolonged delay in therapy at 170 bpm or higher. Both strategies were compared with conventional programming.

During 1.4 years of follow-up, the high rate and delayed therapy programming strategies reduced the risk of a first occurrence of an inappropriate shock by about 80 percent and reduced all-cause mortality by 55 percent.

The study was conducted using transvenous ICDs, but the encouraging results are consistent with the approaches used with subcutaneous ICD technology. Essentially, fewer shocks reduce risk for patients, and the subcutaneous ICD is more reliable at arrhythmia discrimination than are conventionally programmed transvenous ICDs. However, the transvenous ICD technology can further reduce shocks by using anti-tachycardia pacing, which causes no pain, if the ICD is programmed appropriately.

“Recent data collected through the MADIT-RIT trial indicate that it is probably smarter if a device is programmed not to ‘rush’ to shock the heart back into rhythm, and to be sure there is a sustained tachyarrhythmia before treating the rhythm,” Dr. Wilkoff says. “Appropriately programmed transvenous ICDs and subcutaneous ICDs seem to reduce a lot of the mechanisms for inappropriate detection.”

Advertisement

Clear candidates for each ICD type

About half the patients who need an ICD could be candidates for subcutaneous devices, Dr. Wilkoff estimates. The other half require a pacing device, which is possible only with a transvenous ICD. The latter group includes patients with cardiac dyssynchrony and congestive heart failure, who need cardiac resynchronization that requires placement of three leads into the heart.

Patients with venous access problems and those on dialysis, in whom additional leads in the veins should be avoided, are good candidates for subcutaneous ICDs.

Subcutaneous ICDs could be a game changer, Dr. Wilkoff believes, although a final conclusion awaits head-to-head comparison between the subcutaneous and transvenous devices. Improvements to first-generation subcutaneous ICDs are expected, including less-bulky designs.

“We don’t know which is the best approach, but we do know there is a group of patients who are not suitable for venous leads, and those patients clearly will benefit from these developments in implant technology,” says Dr. Wilkoff. “On the other hand, some patients need protection from bradycardia or tachycardia with pacing, and traditional ICDs with intelligent programming will be the best solution for those patients.”

Related Articles

Digital heart symbol
August 24, 2021
Artificial Intelligence Methodology Refines Cardiac Risk Stratification in Cancer Patients

Patient-patient network analysis proves to be fast and clinically intuitive

20-HVI-2034260_impella-heart-illustration_650x450
February 2, 2021
Preemptive Hemodynamic Support for Improving the Safety and Efficacy of VT Ablation

How we’re using a new multidisciplinary approach to broaden the benefits of ablation

20-HVI-2026375_machine-learning_650x450
January 5, 2021
Study Shows Machine Learning’s Potential to Predict Cancer Therapy-Related Cardiac Risk

Models developed with promising accuracy and generalizability to clinical practice

20-HVI-2006254-Image-of-Issue-3D-printing-in-tricuspid-valve-disease-Hero
November 25, 2020
How 3D Printing Can Optimize Surgery Planning and Patient Communication for Primary Tricuspid Valve Disease

Illustrated case series profiles a valuable tool for a rare and complex entity

20-HVI-1961366_home-blood-pressure-monitor-_650x450
September 9, 2020
Digital Remote Blood Pressure Management: Has Its Time Come?

Keys to success include a team-based approach and integration into clinical workflow

20-HVI-1941484_robotic-myectomy-mitral-valve-repair_650x450
August 24, 2020
Introducing Robotic Myectomy With Mitral Valve Repair

A minimally invasive, single-incision approach to two coexisting problems

3D-printed replica of a tricuspid valve with an implanted valved stent
December 30, 2019
3D Printing in Percutaneous Structural Heart Disease Interventions: A Status Report

New review outlines applications to date, hurdles to overcome

19-HRT-5504-Khatri-CQD-2
October 22, 2019
Robotics in the Cardiac Cath Lab

A long-overdue technology is poised to reshape practice

Ad