High-Power, Short-Duration Approach to Atrial Fibrillation Ablation Improves Durability of Results

Strategy also shortens procedural time

atrial fibrillation ablation

A strategy of high-power (50 W), short-duration (8-15 sec) radiofrequency ablation during pulmonary vein isolation (PVI) results in improved long-term PVI durability compared with conventional moderate-power (30-40 W), moderate-duration (20-30 sec) ablation in patients with atrial fibrillation (AF). The high-power, short-duration (HP-SD) approach also shortens the total ablation time.

Advertisement

Cleveland Clinic is a non-profit academic medical center. Advertising on our site helps support our mission. We do not endorse non-Cleveland Clinic products or services. Policy

So concludes a prospective study presented virtually this week during the Annual Scientific Sessions of the Heart Rhythm Society and published simultaneously in JACC: Clinical Electrophysiology.

“The incidence of AF recurrence after ablation is substantial, largely because of a high rate of electrical reconnection between the pulmonary vein and the left atrium,” says the study’s principal investigator, Elad Anter, MD, Associate Section Head of Cardiac Electrophysiology at Cleveland Clinic and Director of the Mark Josephson and Andy Wit Electrophysiology Research Laboratory.

Building on preclinical findings

To improve the clinical outcome of AF ablation and address the challenge of pulmonary vein reconnection, Dr. Anter’s translational research team has studied the biophysics of radiofrequency ablation using advanced computer simulations, ex vivo bench models and preclinical in vivo studies. They found that ablation at higher powers for shorter durations increases the desired effect on target tissue in immediate proximity to the catheter while reducing the undesired effect of ablation on neighboring structures, such as the esophagus and phrenic nerves. This strategy essentially shifts the classic energy distribution to increased heat by resistance and reduced heat by conduction (Figure).

Visualization between high-power, short-duration (HP-SD) and moderate-power, moderate-duration (MP-MD) ablation strategies.

Figure. Visualization of key differences between high-power, short-duration (HP-SD) and moderate-power, moderate-duration (MP-MD) ablation strategies.

The team’s pivotal studies (JACC Clin Electrophysiol. 2018;4:467-479 and J Cardiovasc Electrophysiol.2018;29:1287-1296) have resulted in coining of the term HP-SD ablation and invigorated development of multiple technologies using HP-SD ablation strategies. “However, while this ablation concept has been intriguing, long-term outcome data had not been previously presented,” Dr. Anter observes.

Advertisement

The study in brief

He and his coinvestigators prospectively enrolled 112 patients with symptomatic paroxysmal or persistent AF to undergo AF ablation using the HP-SD ablation strategy — i.e., at a power of 50 W for 8 sec in the posterior wall and 15 sec in the ridge of the left pulmonary vein and the septal aspect of the right pulmonary vein. Results were compared with those of the preceding 112 patients who underwent ablation using a moderate-power, moderate-duration (MP-MD) strategy — i.e., 20 W in the posterior wall and 35-40 W in the in the ridge of the left pulmonary vein and the septal aspect of the right pulmonary vein applied for 20 to 30 sec using a similar catheter technology. Acute pulmonary vein (PV) reconnection was evaluated in all patients and chronic PV reconnection was examined in patients who required redo procedures (n = 18 for HP-SD; n = 23 for MP-MD).

Key findings included the following:

  • Rates of both acute and chronic PV reconnection were significantly lower with the HP-SD ablation strategy compared with the MP-MD strategy — 14/225 (6.2%) vs. 29/231 (12.5%) for acute (P = 0.02), and 16.6% vs. 52.2% for chronic (P = 0.03).
  • The reduced incidence of PV reconnection with the HP-SD ablation strategy resulted from improved catheter stability. Areas of chronic PV reconnections were associated with catheter motion ≥ 1 mm for ≥ 50% application duration. Catheter stability at this high level was more frequently obtained during shorter ablation applications using the HP-SD ablation strategy (88.6% vs. 72.8%; P < 0.001).
  • The HP-SD ablation strategy was less effective for ablation in thicker tissue, particularly for mitral lines. In these locations, ablation using lower powers applied for longer durations was required in order to achieve a transmural effect with low incidence of steam pops.

Advertisement

“Applying higher powers for shorter durations appears effective for PVI, as it reduced the incidence of PV reconnections and the overall procedural duration,” says Dr. Anter. “However, these data cannot be oversimplified. High power should be used diligently with careful attention to changes in impedance, temperature and electrograms. In particular, this strategy is effective for PVI and thin walls but not for thicker myocardium, which requires ablation at lower powers applied for longer durations to achieve transmurality without overheating the subendocardium.”

A key finding of the study, he adds, is the importance of catheter stability for lesion durability. “The accuracy of catheter stability required for creating durable ablation lesions is so high (< 1 motion) and is difficult to achieve when ablation is performed over relatively longer durations,” he says. “The advantage of the HP-SD ablation strategy is that this stringent stability is required only for a short period of time.”

Refinements underway as practice starts to shift

Dr. Anter notes that catheter and generator technologies are now being refined to allow HP-SD ablation in an even safer and more effective fashion. This involves new catheter designs with larger surface areas, multiple temperature sensors and high-energy generators. Dr. Anter’s research team has been involved in the development of these technologies and their evaluation in humans.

“While no single ablation strategy is suited to all patients and all sites, electrophysiologists are starting to move toward ablation at higher power for shorter duration for improved long-term durability of pulmonary vein isolation,” observes Dr. Anter’s colleague, Oussama Wazni, MD, Section Head of Cardiac Electrophysiology at Cleveland Clinic. “The encouraging results from this study should accelerate that shift.”

Advertisement

Related Articles

axial image of the human chest showing the heart and other organs

How Imaging Can Guide Smarter Aortic Valve and Root Decisions

New review proposes a multimodality framework for matching patients to procedures

color cardiac echo next to model of a bulging aorta

Distinctive CME on Valve Imaging and Connective Tissue Disorders on Tap for March 2027

In-depth live symposia in Florida and Cleveland feature interdisciplinary experts from around the nation

yellow and red illustration of blood cells moving through a blood vessel

First-in-Human Data Show One siRNA Can Lower Both LDL-C and Triglycerides

Early results for ARO-DIMER-PA point to a more streamlined way to treat mixed hyperlipidemia

blood vessel on top of a DNA double helix

CRISPR-Cas9 Gene Editing Therapy Safely Slashes Severely Elevated Triglycerides in Early Trial

Findings support further studies of potential one-time treatment strategy

two side-by-side medical illustrations with red trend line overlays

Lipoprotein(a) Reduction With Pelacarsen Has No Impact on Cardiac Events, Lp(a)HORIZON Finds

But unexpected results are ‘not the end of the story’ for Lp(a) lowering, investigators say

stylized heart and lungs with text overlay

Outcome Updates in Adult Cardiac Surgery, Ischemic Heart Disease and Vascular Disease

A scannable snapshot of recent volumes and results

illustration of aortic arch after stent graft repair

Evolving Strategies in Aortic Arch Repair (Podcast)

Earlier detection is reshaping timing and safety of these increasingly common surgeries

illustrated human gut with microbes and an overlay of a rhythm tracing

TMAO Shown to Be a Modifiable Risk Factor for Atrial Fibrillation

Bench-to-bedside study identifies mechanism and potential therapeutic target

Ad